Interventional Equipment
What’s a Heostat?
Forceps
Mini Access Kit (Micropuncture Set)
Used for small vessels or where an especially precise puncture is needed
Puncture needle small caliber 21G seldinger style, and the wire it accepts is a .018-inch wire with
a stiff shaft and a straight floppy tip
o 21G provides precise movement
Includes two part dilator that ultimately will allow for the insertion of an .035-inch wire
o Trying to allow conversion of a small access using at .018 inch wire to a larger guidewire
(0.035 inch wire)
o French- not like gauge (as size goes up, diameter goes up)
o Larger needle over top of the guidewire to stretch the access point
Co-axial Access Set (Neff or Accustick set)
the non vascular version of the mini access set
o drainages
21 or 22G needle (Chiba), long with an inner stylet, accepts .018 wire
o Chiba: long needle, upwards of 15cm
o Need a longer needle that is also more rigid harder to control
find more resources at oneclass.com
find more resources at oneclass.com
Unlock document

This preview shows pages 1-3 of the document.
Unlock all 10 pages and 3 million more documents.

Already have an account? Log in
Two piece dilator ultimately allows for use of .035 wire
Used for percutaneous trans hepatic cholangiography or difficult nephrostomy with an
undilated collecting system
o Lumbar punctures etc
Trocar
employed in drainage procedures
For large pathologies, not located in a critical structure
Larger gauge > ie. 17-19G
Can come simply as a needle, or as part of a drainage catheter
o Parts:
Cutting stylet (inside of:)
what does the actual puncture
Cannula (inside of:)
Catheter (sometimes)
find more resources at oneclass.com
find more resources at oneclass.com
Unlock document

This preview shows pages 1-3 of the document.
Unlock all 10 pages and 3 million more documents.

Already have an account? Log in
Catheter already on top
Stylet and cannula get pulled out and the catheter is in
Ho it’s used:
1. Introduced percutaneously into area of interest (eg. Abscess)
2. Stylet then removed
3. Cannula held in place while catheter is positioned to desired location (may need guide
wire)
4. Contrast medium to confirm position
5. Syringe (for small abscess) or extension tubing and a urinary drainage bag are attached
to drain abscess
o Always try to keep it below the site (even if it is a vacuum container)
Cook catheter (torcon blue biliary manipulation catheter)
Short single-end hole steel braided catheter:
o for urinary & biliary work (deep structures: we need something that graded, metallic
structures woven together)
o excellent torque control (efficient)
Reality Dr. will have a preferred catheter they like to use to navigate
Steal braided inside the blue part so that when the doc turns it, the tip turns as well (torque
control)
Steel-graded eause e do’t at it to kik hard gettig through the apsule of the kide
Biopsies- FNA
Fine needle used for aspiration with a 10ml syringe attached
o Fine Needle Aspiration (taking a few cells at a time, different than core biopsy)
Suction applied to withdraw a few cells for cytology
Often will have lab tech present with a microscope, aspirated cells sprayed on slide and
immediately viewed by lab tech to ensure adequate specimen/ not damaged
find more resources at oneclass.com
find more resources at oneclass.com
Unlock document

This preview shows pages 1-3 of the document.
Unlock all 10 pages and 3 million more documents.

Already have an account? Log in

Document Summary

Trocar: employed in drainage procedures, for large pathologies, not located in a critical structure, can come simply as a needle, or as part of a drainage catheter. Larger gauge > ie. 17-19g: parts, cutting stylet (inside of:, what does the actual puncture, cannula (inside of:, catheter (sometimes, catheter already on top, stylet and cannula get pulled out and the catheter is in, ho(cid:449) it"s used: Biopsies: core sample: used to obtain large core(s) of tissue, cutting needle used, needle often used with an automated gun, http://www. youtube. com/watch?v=hyacp_k_xk4, 1st dx is usually core sample a(cid:374)d the(cid:374) if looki(cid:374)g for (cid:373)ets tha(cid:374) it"s usuall(cid:455) a(cid:374) aspiratio(cid:374) Biopsies: co-axial biopsy needle: the inner portion shoots out and the outer cannula come out, cuts a portion and goes into the inner stylet. Ureteric stent systems: double j configuration, guide wire straightens, nothi(cid:374)g e(cid:454)ter(cid:374)al to (cid:271)od(cid:455).

Get access

Grade+20% off
$8 USD/m$10 USD/m
Billed $96 USD annually
Grade+
Homework Help
Study Guides
Textbook Solutions
Class Notes
Textbook Notes
Booster Class
40 Verified Answers
Class+
$8 USD/m
Billed $96 USD annually
Class+
Homework Help
Study Guides
Textbook Solutions
Class Notes
Textbook Notes
Booster Class
30 Verified Answers